建立了ICP-AES测定高纯Eu2O3中14种稀土元素的方法。用正交设计法优化仪器测量条件,实验了基体Eu2O3对被测元素的干扰和对方法测定下限的影响。14个稀土元素的测定下限为:Yb2O30.5μg·g-1,Y2O31μg·g-1,Dy2O3、Ho2O3、Tm2O3和Nd2O35μg·g-1,Gd2O3、Sm2O3、Er2O3、Ce2O3、Pr6O11和La2O310μg·g-1,Lu2O320μg·g-1以及Tb4O730μg·g-1。加入回收率为88-112%,六次取样分析的相对标准偏差小于11%。方法简便不经化学分离,可用于99.9-99.99%Eu2O3中稀土杂质的测定或生产控制分析。
A method for the determination of 14 rare earth elements in high purity Eu2O3 by ICP-AES was established. The orthogonal design method was used to optimize the measurement conditions of the instrument. The interference of Eu2O3 on the element under test and the influence on the lower limit of the method were tested. The determination limits of 14 rare earth elements are as follows: Yb2O30.5μg · g-1, Y2O31μg · g-1, Dy2O3, Ho2O3, Tm2O3 and Nd2O335μg · g-1, Gd2O3, Sm2O3, Er2O3, Ce2O3, Pr6O11 and La2O310μg · g-1 , Lu2O320μg · g-1, and Tb4O730μg · g-1. The recovery rate was 88-112%. The relative standard deviation of six sampling analyzes was less than 11%. The method is simple and convenient without chemical separation and can be used for the determination of rare earth impurities in 99.9-99.99% Eu2O3 or production control analysis.