目的确立常州地区血栓弹力图(TEG)alpha角值参考范围。方法按照纳入标准搜集363例均进行了TEG普通杯检测和纤维蛋白原(Fib)检测的住院患者的检测值,对alpha角值等数据进行统计分析。结果 1本研究alpha角值参考范围的下限和上限均高于厂家的参考范围。2女性患者的K值(1.62±0.37)min低于男性(1.81±0.41)min、alpha角值(72.32±3.33)°高于男性(70.52±3.98)°(P<0.05)。3女性患者的alpha角值参考范围较男性患者窄。结论应确立适合本地区的包括alpha角值在内的TEG各参数参考范围。
Objective To establish the reference range of alpha angular value of thrombus elasticity map (TEG) in Changzhou area. Methods According to the inclusion criteria, 363 hospitalized patients with TEG common cup test and fibrinogen (Fib) test were collected, and the data of alpha angle value were statistically analyzed. Results 1 The lower and upper limits of the alpha angular range of the study were higher than the manufacturer’s reference range. 2 (1.62 ± 0.37) min in female patients was significantly lower than that in male (1.81 ± 0.41) min, and the alpha value was 72.32 ± 3.33 ° higher than that of male (70.52 ± 3.98) ° (P <0.05). 3 female patients with a narrower alpha angle reference range than men. Conclusion The reference range for each parameter of TEG, including the alpha value, should be established for this region.