根据NEMA标准对一台SPECT进行固有均匀性、旋转中心校正、四象限铅栅模型、Jaszczak系统模型、Hoffman脑3D模型测试。2003–2010年质控数据分析结果显示,有效视野(UFOV)固有积分均匀性与旋转中心均无明显差异(p>0.05),中心视野(CFOV)固有积分均匀性从2007年开始有明显差异(p<0.05)。表明仪器随着年限的延长,性能参数逐渐变差,通过质控能将参数维持在标准范围,保证图像质量以及诊断符合率。
According to the NEMA standard, a SPECT was subjected to inherent homogeneity, rotation center correction, four-quadrant lead grid model, Jaszczak system model and Hoffman brain 3D model test. The analysis of quality control data from 2003 to 2010 showed that there was no significant difference (P> 0.05) between the inherent integral uniformity of the effective field of view (UFOV) and the rotation center, and the inherent integral uniformity of the central field of view (CFOV) p <0.05). It indicates that the performance parameters of the instrument gradually deteriorate with the extension of years, and the parameters can be maintained within the standard range through quality control to ensure the image quality and the coincidence rate of diagnosis.