The Servo 900 ventilator produced by Siemens in Sweden is easy to use, fully functional, and stable. Most of the 900C and 900B ventilator malfunctions occur in the airway section. This article will give an introduction to its common faults and inspection methods. During the inspection, the faceplate parameters are set to standard position (having green color file). Usually the air source alarm or working pressure swing exceeds 170Pa (5cm H2O), the necessary inspections should be made, but the minute ventilation should not be affected. premise. Insufficient air supply: After taking off the air oxygen mixer, touch the gas outlet valve of the mixer block speed joint with a screwdriver, and you can hear the rapid and short blasting sound, indicating that the pressure is enough. On the contrary, slower airflow sounds indicate insufficient air supply or air failure. The reason is usually: