北京医用射线厂产的F76—Ⅲ立式凸轮滤线器摄影架系与该厂产的F78—Ⅲ型医用X线机配套使用的,而我站的X线机为F78—Ⅰ型,不能直接连接使用。我们采取了下述的接线方法,解决了配套使用的问题,且效果良好。图1为F76—Ⅲ型摄影架原理图,图2为F78—Ⅰ型X线机摄影控制部分原理图。将图1CT 1—2与CT1—4用导线直接接至图2X线机继电器JC7线圈01,CT1—1接至
Beijing medical ray plant F76-III vertical cam filter frame and the factory production of F78-Ⅲ type of medical X-ray machine supporting the use of the X-ray machine I station for the F78-Ⅰ type, not directly Connect to use. We have taken the following wiring methods to solve the problem of supporting the use of, and the effect is good. Figure 1 for the F76-Ⅲ-type photographic frame schematic, Figure 2 for the F78-Ⅰ X-ray machine photographic control part of the principle diagram. Figure 1CT 1-2 and CT1-4 with wire directly connected to Figure 2X wire machine relay JC7 coil 01, CT1-1 to