我科1996年购置的CD-1700血液分析仪在近段使用时,曾出现一次“C0untoverrange”,白细胞不能计数的故障.现把我们处理过程简述如下,供同行参考.故障表现:在开机仪器自动进人背景测试后,显示背景良好.进行标本测试时,白细胞不能计数,直方团呈一条抛物线(如图1).屏
When the CD-1700 Blood Analyzer purchased by our department in 1996 was used in the near segment, there was a “C0untoverrange” and the white blood cells could not count the failure. Now we briefly describe the process as follows for your reference. Failure performance: After the boot instrument automatically enters the background test, the display background is good. When the specimens were tested, white blood cells could not be counted, and the histogram showed a parabola (Figure 1). Screen