论文部分内容阅读
LH2、LN2为冷源.用减压降温和增压升温的方法.分别获得了14K~30K、63K~100K的浴槽温度。恒温精度优于±50mK/20min。LH2耗量<2.5/h。LN2耗量<0.6L/h。适用于热电偶温度计和电阻温度计的比对校准。
LH2, LN2 cold source. With decompression cooling and heating method. Respectively obtained 14K ~ 30K, 63K ~ 100K bath temperature. Temperature accuracy is better than ± 50mK / 20min. LH2 consumption <2.5 / h. LN2 consumption <0.6L / h. Suitable for thermocouple thermometer and resistance thermometer calibration.