根据近两年到所辖各医疗单位现场检定台式血压计共1858台的情况 ,对其中584台不合格的原因进行了统计 ,结果如下 :不合格原因分析及故障的排除 :(1)气密性差 :前气阀开关应用时间过长造成磨损漏气 ;袖带中橡胶带和充气球回气阀、橡皮球老化造成漏气等。排除方法 :更
According to the past two years to the medical units under the jurisdiction of the scene a total of 1858 sets of desktop sphygmomanometer conditions, of which 584 unqualified reasons for the statistics, the results are as follows: Analysis of unqualified causes and troubleshooting: (1) Airtight Poor sex: the valve switch is too long before the application of wear and tear caused by leakage; cuff in the rubber band and inflatable ball valve, rubber ball aging caused by leakage and so on. Remedy: more