故障现象:无稀释剂进入白细胞池;溶血剂泵无液而有大量气泡出现。分析与处理:这种故障应从下面几方面着手分析处理:(1)检查液路。首先停机检查稀释剂本身,再检查输液管,用注射器逐级抽吸稀释液,直至在终端口见到液为止,这就可证明液路有无问题了。在停机检查时电磁阀
Symptoms: no diluent into the white blood cell pool; hemolysis pump without liquid and a large number of bubbles appear. Analysis and processing: This failure should be analyzed and dealt with the following aspects: (1) check the liquid path. First shut down to check the diluent itself, and then check the infusion tube, the syringe with a gradual aspiration of diluent until the end of the mouth to see the liquid so far, which can prove that there is no problem with the liquid. Solenoid valve during shutdown check