在数模变换器(以下简称D/A)线性误差自动测试系统中,为使数字电压表(以下简称DVM)SM215MK2正常工作,在相邻两次启动DVM之间,控制程序必需插入约0.7秒的延时循环,从而显著地降低了系统的测试速度。考虑到D/A需要作大量的测试,当位数增加时,测试点更是成倍增长,因此有必要设法提高该系统的测试速度。本文将从DVM本身、计算机的接口和程序设计三个方面,提出改进措施,使系统的测试速度可提高一倍以上,约为2次/秒。文中对DVM的改进,实际上也可适用于其他双斜率积分型DVM。
In the D / A linearity error automatic test system, for the digital voltmeter (hereinafter referred to as DVM) SM215MK2 to work normally, the control program must be inserted between two adjacent DVMs for about 0.7 seconds Of the delay cycle, which significantly reduces the system’s test speed. Taking into account the D / A requires a large number of tests, when the median increases, the test point is doubled, it is necessary to try to improve the system test speed. This article will propose improvements from three aspects of the DVM itself, the computer interface and program design, so that the test speed of the system can be more than doubled, about 2 times / sec. The improvement of the DVM in this paper can be applied to other dual slope integral DVMs.