论文部分内容阅读
故障现象: 低电导报警,状态红灯亮,偏路开关进行偏路。检查步骤与方法: 1.低电导报警时首先要看一下浓缩液,如果浓缩液是同一批次的就要对透析液进行检查,看是否正常。因为此系列透析机,透析液的配比是固定比例的比例泵(水:浓缩液=34:1)如果比例泵出了故障及孔堵塞等都会使配比后的透析液偏离正常值,引起电导率报警。进行对透析液化验的目的,是区分报警故障引起的原因,是比例泵还是电导率检测系统。我们在取样口取透析液经化验钠含
Symptoms: low conductivity alarm, the status of red light, partial circuit breaker bias. Check the steps and methods: 1. Low conductivity alarm first look at the concentrate, if the concentrate is the same batch of the dialyzer to check to see if it is normal. Because of this series dialysis machine, dialysate ratio is a fixed ratio of the proportion of the pump (water: concentrate = 34: 1) If the ratio of pump out of the hole and plug the proportion will be so deviated from the normal dialysate, causing Conductivity alarm. The purpose of conducting a dialysis fluid test is to distinguish the cause of an alarm failure, a proportional pump or a conductivity detection system. We take dialysate in the sampling port by the test sodium