目前,半自动生化分析仪的使用已普及,在使用过程中难免出现一些小故障,影响仪器的正常工作及结果的准确性。现将我们使用的荷兰威图ISP—2半自动生化分析仪故障排除的体会报道如下。1零点不稳定1.1主要表现多次试剂空白调零,零点显示差异大。1.2原因比色皿有污物沉积或杂
At present, the use of semi-automatic biochemical analyzers has become widespread. In the course of use, it is inevitable that some small faults will occur, affecting the normal operation of the instrument and the accuracy of the results. Now we are using the report on the troubleshooting of the Rittal ISP-2 semi-automatic biochemical analyzer. 1 Zero instability 1.1 The main performance of multiple reagent blanks zero, zero shows a large difference. 1.2 Causes There are dirt deposits or impurities in the cuvette