故障现象:定位时突然出现监视器无图像mA表无指示,透视kV表指示小于10kV。检查分析:电源电压指示220V,透视kV小于10kV,调查透视kV调节器旋钮,表针指示不动;踏下脚闸,透视继电器工作,mA表无指示。根据现象和线路分析初步判断:(1)透视继电器
Symptoms: There is no image on the monitor when positioning. There is no indication on the mA meter. The perspective of the kV meter is less than 10kV. Inspection and analysis: power supply voltage indication 220V, perspective kV less than 10kV, investigation perspective kV regulator knob, watch hands do not move; step foot gate, see through relay work, mA watch no indication. According to the phenomenon and the line analysis preliminary judgment: (1) perspective relay