关于高压消毒时制剂瓶炸毁原因和预防办法,我院从1961年起即做了初步探讨,当时是用热蒸馏水,将滤液温度由10-15℃提高到30-35℃的办法,炸瓶率由5%降至1%。灭菌时液体温度平均升高速度为4℃/分。但因滤液温度升高,使玻璃膨胀,滤器的滤孔增大,液体的澄明度下降,因此1962年10月改用如下方法后,进一步降低制剂瓶炸毁率,也提高了液体的澄明度。(一)方法:先开高压消毒器的外鍋(夹层鍋),以1公斤/方平厘米的压力预热30-40分钟。此刻液体温度由10-15℃升至30℃以上。然后闭上外鍋,打开内鍋进气栓,按常法进行灭菌。灭菌时液温升高速度亦平均为4℃/分左右。
On the high-pressure disinfection bottle blow-up causes and prevention measures, our hospital since 1961 to do a preliminary study, was the time with hot distilled water, the filtrate temperature from 10-15 ℃ to 30-35 ℃ approach, deep-frying The rate dropped from 5% to 1%. The average temperature rise of the liquid during sterilization is 4 ° C / min. However, due to the filtrate temperature increases, the expansion of the glass, the filter hole increases, the clarity of the liquid decreased, so October 1962 switch to the following method, to further reduce the bottle blown rate, but also improve the clarity of the liquid . (A) Method: First open the outer pot of high-pressure sterilizer (sandwich pot) to 1 kg / square centimeter pressure preheat 30-40 minutes. At the moment the liquid temperature rose from 10-15 ℃ to 30 ℃ above. Then close the outer pot, open the inner air intake plug, according to the usual method of sterilization. Sterilization liquid temperature increase rate is also an average of 4 ℃ / min or so.