论文部分内容阅读
采用VA-90气态原子化装置与WFX-1D原子吸收分光光度计联用测定饮料中铅,采用HCl-NaBH4-K3Fe(CN)6体系,经L9(34)正交试验选择最佳实验条件为:NaBH420g/L,K3Fe(CN)6100g/L,HCl1.5ml,氩气流量为0.8L/min。在0~100μg/L范围内,相关系数r=0.9996,回收率为94.1%~107.5%,R·S·D%=3.3~5.2特征浓度和检出限分别为1.29μg/L/1%和0.049μg/L。干扰试验表明,Sn、Hg、Cr、Cu、Mn在500μg/L以下,Zn、As、Fe、Ni、Ag在1000μg/L以下,不干扰测定
The determination of lead in beverages by VA-90 gas atomization apparatus and WFX-1D atomic absorption spectrophotometer was used to determine the optimum conditions for the determination of lead in the beverage by HCl-NaBH4-K3Fe (CN) 6 system. : NaBH420g / L, K3Fe (CN) 6100g / L, HCl1.5ml, argon flow 0.8L / min. In the range of 0-100 μg / L, the correlation coefficient r = 0.9996, the recoveries were 94.1% -107.5%, and the R · S · D% was 3.3-5.2. Respectively 1.29μg / L / 1% and 0.049μg / L. Interference tests showed that the content of Zn, As, Fe, Ni and Ag was less than 500μg / L for Sn, Hg, Cr, Cu, Mn and did not interfere with the determination