对氯苯氧乙酸在流速为 1ml/min ,淋洗液为 0 .8mmol/l的Na2 CO3和 0 .5mmol/l的NaOH的混合溶液的离子色谱仪DIONEX(DX -1 0 0T)上较好的检测。其检测限为 0 .1 8μg/ml(S/N =2 .5) ;保留时间、峰高、峰面积的变异系数分别为 1 .0 967%、0 .9786%、0 .583 6% ,在线性范围 3 0 -3 0 0 μg/ml,峰高、峰面积的线性回归系数分别为 0 .9977和 0 .9971。在此条件下 ,常规阴离子对对氯苯氧乙酸的测定没有干扰。对土壤样品进行测定 ,回收率分别为 99.70 % (峰高 )。这种方法快速、灵敏 ,结果令人满意。
P-Chlorophenoxyacetic acid is better on ion chromatograph DIONEX (DX-100T) with a mixed solution of a flow rate of 1 ml / min, an eluent of 0.8 mmol / l Na2CO3 and a 0.5 mmol / l NaOH The test. The detection limit was 0.18μg / ml (S / N = 2.5). The coefficients of variation of retention time, peak height and peak area were 1. 967%, 0 9786% and 0 5 583% In the linear range of 3 0 -3 0 0 μg / ml, the linear regression coefficients of peak height and peak area were 0.9977 and 0.9971, respectively. Under this condition, the conventional anion pair did not interfere with the determination of p-chlorophenoxyacetic acid. Soil samples were measured, the recovery rates were 99.70% (peak height). This method is fast, sensitive and the result is satisfactory.