【摘 要】
:
该机常见故障是开机后ABS值出现E9999,一般从以下几方面检查 :①流动池脏或有气泡 :②光源灯不亮或灯泡老化光强不够 ;③滤光片脏。通过以上检查、处理 ,故障一般可以排除。
【机 构】
:
第三军医大学西南医院,第三军医大学新桥医院,第三军医大学新桥医院 重庆400038
论文部分内容阅读
该机常见故障是开机后ABS值出现E9999,一般从以下几方面检查 :①流动池脏或有气泡 :②光源灯不亮或灯泡老化光强不够 ;③滤光片脏。通过以上检查、处理 ,故障一般可以排除。一、故障现象 :开机后ABS值为 2 76 ,但在按“ZERO”自动调零后出现“E9999”。二、分析及排除 :该机自
The common fault of this machine is that E9999 appears when the ABS value is turned on. Normally, check from the following aspects: 1 The flow pool is dirty or has air bubbles: 2 The light source lamp is off or the light bulb is not strong enough; 3 The filter is dirty. Through the above inspections and handling, faults can generally be eliminated. First, the phenomenon of failure: After the boot ABS value of 2 76, but after pressing “ZERO” automatically after the zero “E9999”. Second, analysis and exclusion: The machine