论文部分内容阅读
本文报告应用国产WFX—1A型原子吸收分光光度计测定山羊瘤胃液和血清的钙和镁,并进行实验讨论。血液钙的变异系数和回收率分别为0.67%知102%,镁的变异系数和回收率分别为0.55%和104%。实验结果与常规法比较差异不显著(P>0.05)。
This paper reports the determination of calcium and magnesium in the rumen fluid and serum of goats using a domestic WFX-1A atomic absorption spectrophotometer and discusses the experimental results. The coefficient of variation and recoveries of blood calcium were 0.67% and 102%, respectively. The coefficient of variation and recovery of magnesium were 0.55% and 104%, respectively. The experimental results showed no significant difference compared with the conventional method (P> 0.05).