论文部分内容阅读
Cholesterol oxidase biosensor has been constructed by using bovine serum albumin and glutaraldehyde as cross linker to immobilize cholesterol oxidase and cholesterol esterase on a glassy carbon electrode modified with Nafion and methyl viologen. The biosensor has been used to determine total cholesterol in blood. The linear range of the determination is 2.5×10~7 to 1.0×10-4 mol/L. The detection limit is about 5.0×10~8 mol/L. The response time is 12 s. This biosensor has the advantage of high selectivity, sensitivity and short response time.
Cholesterol oxidase biosensor has been constructed by using bovine serum albumin and glutaraldehyde as cross linker to immobilize cholesterol oxidase and cholesterol esterase on a glassy carbon electrode modified with Nafion and methyl viologen. The biosensor has been used to determine total cholesterol in blood. The linear range The detection limit is about 5.0 × 10 -8 mol / L. The response time is 12 s. This biosensor has the advantage of high selectivity, sensitivity and short response time.